5 system information, System, Information – K-Patents SeedMaster 2 User Manual
Page 72

10 DISPLAY
72
72
WARNSD
STANDARD DISPLAY P1:95.2%
1 / 0 . 1
SUPERSATURATION :
MASS. SOLIDS C.: %
0.81 72.8
SEEDED
STANDARD DISPLAY P2:92.32%
2 / 0 . 2
SUPERSATURATION :
MASS. DENSITY: kg/m3
1.12 1482
BACK
Fig. 10.8
This is a display – only presentation of the selected data. Information on seeding and on the purity of the feed
syrup currently in use is also presented. The only operation (BACK) results in the return to the previous step
on the selection tree.
10.5 SYSTEM INFORMATION
SYSTEM INFORMATION provides a brief summary on data inputs and outputs and on the selected mode of
seeding (Fig. 10.9).
WARNSD
SYSTEM INFORMATION
1 / 0 . 1
INPUT
OUTPUT
SEEDING
CONC . :
SER .
SER . : 0
AUTO SUPS .
TEMP . :
SER .
0 - 20 : 0
SETP . : 1 . 12
MOTOR:
4 - 20
4 - 20 : 2
SW. ver sion: 1.1
SYSTEM INFORMATION
2 / 0 . 2
INPUT
OUTPUT
SEEDING
CONC . :
SER.
SER . : 0
TEMP . :
SER.
0 - 20 : 0
COMM.
DENS.:
COMM.
4 - 20 : 2
SW. ver sion: 1.1
BACK
Fig. 10.9
The 1
st
row shows :
1.
current information on seeding :
BLANK : blank (no seeding in close as yet)
WARNSD : warning on approaching seeding (see Ch. 7.5)
SEEDED : seeding is over.
2.
INSTRUMENT NO. over TAG.
The INPUT column lists the types of inputs configured for the CONCENTRATION, TEMPERATURE and
THIRD INPUT data. The last one is identified by the actual name of the type of input (MOTOR, DENS., MA.
SOL., CR. CONT. meaning motor consumption, density, massecuite solids content and crystal content
determined by the laboratory) selected during configuration.
Possibilities:
SER.
: Data input directly from the refractometer using communication (COMX port).
4 – 20
: Standard current, mA.
0 – 20
Standard current, mA
MAN
: Laboratory crystal content data (see Ch. 7.3.3), manual entry.
COMM.
: Data transfer via digital communication (COMX or Ethernet).