K-Patents SeedMaster 2 User Manual
Page 23

4
TECHNICAL DATA, MOUNTING AND ELECTRICAL CONNECTIONS
23
23
Calculated outputs
1. Analog (0-20, 4-20 mA) outputs:
Any 2 of the 7 calculated massecuite parameters (per pan).
2. Digital (ON / OFF) outputs:
1. DO1: Warning on approaching seeding (programmable).
2. DO2: Opening the seeding valve for a selected time interval.
NOTE:
All calculated and measured data can be also transmitted via digital communication.
Process interface
Inputs
1. Analog current
8 channels
Symmetrical, galvanically isolated
Range
0(4)-20 mA, keyboard selectable
Input impedance
100 Ohm
Common mode voltage
50 V max.
2. RTD (Pt 100) (option)
2 channels
4 wire connecting mode
3. Digital (ON / OFF)
8 channels
Signal sources:
passive, contact or open collector
active, + 24 V
dc
automatic signal detection
Outputs
1. Analog current
4 channels
Galvanically isolated
Range
0(4)-20 mA, keyboard selectable
Max. load
600 Ohm
2. Digital ON / OFF)
4 channels
Isolated open collector,
overvoltage and short circuit protected
3. Power supply for transmitters
1 X 24 Vdc, 200 mA max.
4. Device Status (ON/OFF)
Operating / Error
Communication
1. Serial (COM1, COM2, COM3)
3 ports
Galvanically isolated
Standards
RS232, RS422, RS485
(keyboard selectable)
Control signals
CTS, RTS
(keyboard selectable)
Baud rates
1200…38400 Baud
(keyboard selectable)
Length of cable
RS232
15 m max.
RS422 /485
1200 m max.
Protocols
K-PATENTS refractometer protocol
MODBUS SLAVE (ASCII, RTU)
2. Ethernet 10 6 100 Base T
Connector
RJ45
Protocols
TCP / IP, MODBUS TCP,
UDP / IP (K-PATENTS refractometer protocol)
Front panel
Display 5,7 ‘ QVGA 320x240 graphic LCD
Keyboard
Membrane switches with flexible foil cover
LED indicators
Power, Run, Alarm
Power supply
100…240 V ac, 50 / 60 Hz; 10-25 VA
24 (18…30) Vdc