Connector interface pin definitions – Fluke Biomedical DataSim 6100 User Manual
Page 57

Operation
Connector Interface Pin Definitions
2
2-21
150
Ω
1
%
W
hite
W
hite
G
N
D Ref
+ SIG
+ EXC
- SIG
Ref Pin 4
Pin 2
C Pin 5
G
N
D Pin 6
Black
Pair
fdg004.eps
Figure 2-2. Pressure Cable Wiring for Monitor with 40 µV/V mmHg
Sensitivity
Note
To minimize noise, the +/- SIG wires should be a twisted pair, as
should the +/- EXC wires
Connector Interface Pin Definitions
Table 2-8 contains definitions for ECG cable connector interface pins.
Table 2-9 contains definitions for pressure cable connector interface pins.
Figure 2-3 shows plug connector pin assignments.
Table 2-8. ECG Connectors
Pin #*
Function
1 N/C
2 N/C
3
High Level Out
4 +9
V
5 Defib
In
6 GND
7
Pacer Flag (5V, 10 msec)
See also other documents in the category Fluke Biomedical Equipment:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)