Interactive iabp personality modules – Fluke Biomedical DataSim 6100 User Manual
Page 52

DataSim 6100
Operators Manual
2-16
Table 2-6. Personality Modules (cont.)
MDE Ref #
Part #
Description
Additional Detail
6070-29B 2244678
Interactive 
IABP/Datascope 
System 90
6070-30B 2244684
Interactive 
IABP/Aries 
Early Inflation 
Late Inflation 
Early Deflation 
Late Deflation 
Proper Timing 
6070-37B 2244706
Normal/Disease 
Heart 
Interactive IABP Personality Modules
The Interactive Intra-Aortic Balloon Pump Personality Modules are used with 
the Simulator and Kontron, Datascope System 80, Datascope System 90, and 
Aries monitors. Invasive arterial BP and sync timing cables are included with 
each module. 
To generate the special waveforms:
1. Determine the need for a balloon according to the monitor used.
•
Datascope 90: It is essential that a balloon (or suitable replacement) 
be connected to the 
Balloon Output
connector for proper operation
of the Datascope 90 pump.
•
Datascope 80, Kontron Aries: It may be necessary to defeat the alarm 
by connecting a balloon or suitable replacement to the 
Balloon
Output
connector.
2. Connect Datascope/Kontron/Aries ECG cable to the Simulator
ECG
snap
connectors.
3. Insert the Interactive IABP module into the Simulator
Personality
Module
connector.
4. Power on the Simulator.
5. Connect the following cables to their respective monitors: