Table 1-5. Pendant Keypad Codes and Functions (cont.)
Code Function
Description
34 SENSE
and
CAPTURE FAIL
Frequent loss of capture, low amplitude 
escape beat, sensing failure of escape 
beat 
35 FAILURE
TO
CAPTURE
No capture; pacer artifact amplitude is 8 
mV with 1 msec duration. 
MISC ECG
38 CARDIAC
FAILURE
Intermittent PVCs and couplet followed 
by ventricular tachycardia, followed by 
ventricular fibrillation, followed by 
asystole 
39
ST ELEVATION
Regular rhythm with ST Elevation at 2 
mm 80 msec after J point 
40
ST DEPRESSION
Regular rhythm with ST Depression at 3 
mm 80 msec after J point 
41
AGONAL
No P waves; wide complexes with 
periods of asystole 
42 VENTRICULAR
ASYSTOLE
No QRS, regular atrial rhythm
43
CONVERSION
Vfib with CPR followed by Vfib without 
CPR for 2 seconds prior to defib: Defib 
artifact, Brady rhythm with CPR, Brady 
rhythm without CPR, followed by NSR 
ECG ARTIFACT
46
DEFIBRILLATION
Positive and negative baseline 
saturation; artifact is automatically 
followed by pre-selected rhythm (see 
#59 below). 
47 60
CYCLE
ARTIFACT
Artifact amplitudes 0.1, 0.4, and 0.7 mV 
p-p; use 
]
and
[
keys or eliminate with
STD key.
48 ECG/RESP
ARTIFACT
Artifact amplitudes 0.4, 0.6 and 1.0 mV 
peak to peak; use 
]
and
[
keys or
eliminate with STD key.