Fluke Biomedical DataSim 6100 User Manual
Page 26

DataSim 6100
Operators Manual
1-14
Table 1-5. Pendant Keypad Codes and Functions (cont.)
Code Function
Description
34 SENSE
and
CAPTURE FAIL
Frequent loss of capture, low amplitude
escape beat, sensing failure of escape
beat
35 FAILURE
TO
CAPTURE
No capture; pacer artifact amplitude is 8
mV with 1 msec duration.
MISC ECG
38 CARDIAC
FAILURE
Intermittent PVCs and couplet followed
by ventricular tachycardia, followed by
ventricular fibrillation, followed by
asystole
39
ST ELEVATION
Regular rhythm with ST Elevation at 2
mm 80 msec after J point
40
ST DEPRESSION
Regular rhythm with ST Depression at 3
mm 80 msec after J point
41
AGONAL
No P waves; wide complexes with
periods of asystole
42 VENTRICULAR
ASYSTOLE
No QRS, regular atrial rhythm
43
CONVERSION
Vfib with CPR followed by Vfib without
CPR for 2 seconds prior to defib: Defib
artifact, Brady rhythm with CPR, Brady
rhythm without CPR, followed by NSR
ECG ARTIFACT
46
DEFIBRILLATION
Positive and negative baseline
saturation; artifact is automatically
followed by pre-selected rhythm (see
#59 below).
47 60
CYCLE
ARTIFACT
Artifact amplitudes 0.1, 0.4, and 0.7 mV
p-p; use
]
and
[
keys or eliminate with
STD key.
48 ECG/RESP
ARTIFACT
Artifact amplitudes 0.4, 0.6 and 1.0 mV
peak to peak; use
]
and
[
keys or
eliminate with STD key.