Heart rate adjustment – Fluke Biomedical DataSim 6100 User Manual
Page 42

DataSim 6100
Operators Manual
2-6
Table 2-3. Calibration Functions (cont.)
Function
Adjusment
Levels
(78) HR CAL
(BPM)
(82) ECG/SIZE/CAL
(mV)
(84) RESP SIZE
CAL (
Ω
)
+8 203
+9 227
+10 250
+11 268
+12 300
To adjust calibration levels:
1. Using the keypad, press
SPEC FUNC
, followed by the numbers
78, 82,
or 84
. For example, using the number
78
, the keypad briefly displays:
ECG RATE
followed by the prompt:
], [ or STD
2. Press ] to increase or [
to decrease the value, or press
STD
to reset to the
standard level.
Heart Rate Adjustment
The Rate Adjust function, Special Function 53, offers the user the option of
adjusting rates up or down in 10% increments. Table 2-4. lists the heart rates
available for the various rhythms. On power-up, each rhythm defaults to the
standard heart rate (STD).
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)