Fluke Biomedical DataSim 6100 User Manual
Page 23

Introduction and Specifications
Instrument
Familiarity
1
1-11
27
. Functions can be selected in any order to simulate virtually any patient
condition sequence.
After the code is input, the following prompt displays:
], [ or STD
At this prompt, the user can adjust the value of the parameter as desired:
•
]
- Increases parameter value; momentarily displays at max“ when
maximum value is reached
•
[
- Decreases parameter value; momentarily displays at min when
minimum value is reached
•
STD - Resets parameter value to the standard value
Some of the Special Functions are actually programmed sequences of rhythms 
or events. Table 1-5 describes the codes and functions. 
Table 1-5. Pendant Keypad Codes and Functions
Code Function
Description
00
NORMAL SINUS
Provides a convenient way to return to 
normal sinus rhythm at a rate of 70 BPM 
01 SINUS
TACHYCARDIA
Rapid regular rhythm with normal P 
wave and heart rate of 140 BPM 
02 SINUS
BRADYCARDIA
Slow regular rhythm with normal P wave 
and heart rate of 35 BPM 
03
ASYSTOLE
No beats but very small irregular signal 
remains 
04 ATRIAL
FLUTTER
Regular ventricular rhythm; atrial flutter 
with 4:1 conduction 
05 ATRIAL
FIBRILLATION
Rapid irregular atrial signal with no real 
P waves; irregular ventricular rate 
06 2nd
DEGREE
AV
BLOCK I
Wenckebach; irregular rhythm with 
normal P waves; PR interval lengthens 
progressively until a dropped beat 
occurs and cycle repeats. 