Artifact simulation – Fluke Biomedical DataSim 6100 User Manual
Page 43

Operation
Artifact
Simulation
2
2-7
Table 2-4. Heart Rates Obtained via Rate Adjust Special Function
RHYTHM -40%
-30%
-20%
-10%
STD
+10%
+20%
+30%
Bradycardia 21 24 28 31 35 38 42 45
Ventricular 21 24 28 31 35 38 42 45
Junctional 24 28 32 36 40 44 48 52
Normal
Sinus
42 49 56 63 70 77 84 91
Sinus Tach
84
98
112
126
140
154
168
182
Atrial Tach
90
105
120
135
150
165
180
195
Atrial
Fib
60 70 80 90
100
110 120 130
Note
For performance check applications, the Heart Rate Calibration,
Special Function 78 (ECG RATE), provides calibrated heart rates
from 30 to 300 BPM, See Table 2-3.
To adjust the heart rate:
1. Using the keypad, press
SPEC FUNC
, followed by the number
53
. The
keypad briefly displays:
RATE
followed by the prompt:
], [ or STD
2. Press ] to increase or [
to decrease the value, or press
STD
to reset to the
standard level.
Artifact Simulation
Various types of artifact can be superimposed on the ECG. The type and level
of artifact is selected via the Special Functions 47, 48, and 49.