Fluke Biomedical DataSim 6100 User Manual
Page 56

DataSim 6100
Operators Manual
2-20
Table 2-7. Appropriate Pressure Simulator Interface Cables (cont.)
Model Item Model
Item
Spacelabs
(400,500,600,PC)
4100-25
Nihon/Kohden (5M)
Physiocontrol (6M)
4100-37
Honeywell/Meddars (9M)
MDE Escort (6M)
4100-40
Fukuda Denshi (8M)
Mennen (6M)
4100-42
Siemens Mingo 7 (15M)
4100-09
IVY (6M)
4100-11 EforM/Honeywell
(6F) 4100-50
Armstrong High Level Output
4100-12 BD/Datamedix/Ohio
(9M)
4100-60
Marquette Eagle Series
4100-61
Fukuda Denshi (12M)
Figures 2-1 and 2-2 show the proper wiring for cables connecting channels 2
through 4 to most pressure monitors. For assistance in wiring pressure cables,
contact the Fluke Biomedical Service Center.
Note
A special interface cable is not necessary for either ECG of
respiration, because the Simulator ECG snap outputs are directly
compatible with most standard patient cable lead wires.
150
Ω
1
%
W
hite
W
hite
G
N
D Ref
+ SIG
+ EXC
- SIG
Black
Pair
fdg003.eps
Figure 2-1. Pressure Cable Wiring for Monitor with 5 µV/V mmHg
Sensitivity