Fluke Biomedical DataSim 6100 User Manual
Page 28

DataSim 6100
Operators Manual
1-16
Table 1-5. Pendant Keypad Codes and Functions (cont.)
Code Function
Description
77 PEDIATRIC
APNEA/BRADY
Intermittent apnea periods; heart rate
drops from 140 BPM to 70 BPM each
apnea period.
CALIBRATION
78 HEART
RATE
CALIBRATION
Calibrated heart rates from 30 to 300
BPM
81 LINEARITY/SPEED
2.5
Hz
triangular ECG waveform
82 ECG
SIZE
CALIBRATION
Cal pulse amplitudes from .25 mV to
2.5 mV
83 ARRHYTHMIA
SEQUENCE
Automatic sequence of arrhythmias
over 3 minute period of time: NSR,
Dropped Beat, AV Pace, V Pace,
Atrial Tach, Couplet, RBBB, Triplet,
Bigeminy (3 Foci), Vtach
84 RESPIRATION
SIZE CAL
Respiratory depth ranging from 0.25 to
2.5 ohm; calibrated for RA-LA lead
BLOOD PRESSURE
62 ZERO
PRESSURE
Zeros all three-pressure channels as
well as auxiliary CO2 channel
63
PATENT LINE
Outputs a 50 mmHg flush waveform on
PA channel
PA waveform follows flush. Trailing edge
of flush and PA pressure waveforms
demonstrate ideal frequency
characteristics of pressure line.
64
RESONANT LINE
Outputs a 50 mmHg flush waveform on
PA channel
PA waveform follows flush. Trailing edge
of flush and PA pressure waveforms
demonstrate resonant frequency
characteristics of pressure line.