Specifications – Fluke Biomedical DataSim 6100 User Manual
Page 31

Introduction and Specifications
Specifications
1
1-19
Specifications
Dimensions
........................................................... 13 in H x 10 in W x 4.7 in D
(33 cm H x 25.4 cm W x 11.94 cm D)
Weight
.................................................................... .05 lb (3.2 kg)
Power
AC power ............................................................ 100 V/240 V, 1.6 A max, 50-60 Hz;
output: +20V, 2.5 A; output power:
50 W max
Battery Type ....................................................... 12 V, 1.9 AH sealed lead acid
rechargeable
Battery Capacity ................................................. 20 hr
Battery Charge Time........................................... 5 to 95 % of complete charge in 14 hr
Temperature ..........................................................
10 ºC to 40 ºC
Humidity ................................................................
80% maximum
ECG
Output signals..................................................... High Level: 1 V/1 mV
Heart Rate
Range ................................................................. 30 to 300 BPM
Accuracy ............................................................. ±1 %
Output Connectors
Low Level............................................................ 12-Lead electrode snaps
High Level........................................................... Switchcraft 15GM7F
Pacemaker Artifact ............................................. -8 mV, 1 ms
Sample Rate (Maximum) .................................... 250 sample/s
Performance Testing
Linearity .............................................................. 2.5 Hz triangular wave
HR Cal Check ..................................................... 30 to 300 BPM
Chart Speed........................................................ 2.5 Hz square wave
Amplitude............................................................ 0.25, 0.50, 1.00, 1.50, 2.00, and
2.50 mV
Blood Pressure
Output signals
High Level........................................................... 1 V/100 mmHg
Transducer.......................................................... 5 and 40 µV/V mmHg
Exciter Voltage.................................................... 10 V ac or dc max
Static Range ....................................................... 0 to 250 mmHg, adjustable in 5 %
increments
Accuracy ............................................................. ±5 % or 1 mmHg, whichever is greater