Fluke Biomedical DataSim 6100 User Manual
Page 19

Introduction and Specifications
Instrument
Familiarity
1
1-7
Table 1-2. Simulator Front Panel Controls and Indicators
Label Component
Description
A
Power Indicator
Illuminates when power is ON
B
Power Switch
Switches power on and off
This battery operated unit is 
rechargeable with the supplied charger. 
C
ECG Snap 
Connectors 
Outputs low level 12 lead ECG
Attach standard patient ECG cable lead 
wires to these snap connectors to 
display the simulated ECG waveforms 
on the monitor screen. 
To minimize 60-cycle artifact, use a 
reference electrode connection. For 
example, most five-lead ECG diagnostic 
recorders require that a Simulator REF 
signal is connected to the RL input of 
the patient cable. The Simulator can 
also output the ECG in a modified chest 
lead configuration (MCL2) used for 
bedside monitoring. Simply attach the 
LL (+) red, RA (-) white, and LA (ref) 
patient cable snaps to the patient 
Simulator. Refer to the operators 
manual of the ECG monitor for hook-up 
instructions. 
D
Pressure Output 
Connectors 
Output for arterial, PA, and RA 
pressures 
This output simulates the electrical 
output of the BP transducer that would 
be used with the patient monitor. The 
Simulator generates signals that are 
compatible with either 5 or 40-microvolt 
transducers. 
