Operation
Defibrillation
Training
2
2-9
followed by the prompt:
CHAN?
2. Enter number
6
. The following prompt displays:
], [ or STD
Press
] t
o increase,
[
to decrease, or
STD
to reset to standard level.
Respiration amplitudes differ, depending upon which lead is monitored. Tables 
2-2 and 2-3 list amplitudes for the RA-LA lead. 
Note
For calibration applications, use the Respiration Size Cal function, 
Special Function 84, to display impedance amplitudes. Pressure 
waveforms are modulated by the respiration waveforms to simulate 
respiration artifact. The respiration artifact level can be varied using 
Pressure/Respiration Artifact, Special Function 67. 
Cardiovascular artifact or ECG/Respiration coincidence can be 
simulated by selecting the CVA Coincidence, Special Function 76. 
The impedance signal is synchronized with the ECG. Impedance 
decreases after each R wave, simulating a decrease in impedance due 
to an increase in pulmonary blood volume during systole. 
Defibrillation Training
The Defibrillation Training mode lets the user interface the Simulator with the 
Laerdal Arrhythmia Anne and the Armstrong Chris Clean manikins for 
realistic defibrillation and ECG paddle monitoring training. 
To train in defibrillation using a manikin:
1. Follow the guidelines provided by the training manikin for proper cable
connection.
2. Plug the torso skin cable into the
DEFIB MANIKIN
jack.
3. Power on the Simulator.
4. On the keypad, select a life-threatening arrhythmia, such as
V-FIB