Fluke Biomedical DataSim 6100 User Manual
Page 27

Introduction and Specifications
Instrument
Familiarity
1
1-15
Table 1-5. Pendant Keypad Codes and Functions (cont.)
Code Function
Description
49 MUSCLE
ARTIFACT
Artifact amplitudes 0.4, 0.6 and 1.0 mV
peak to peak; use
]
and
[
keys or
eliminate with STD key.
50
CPR
CPR artifact a 70/min
ADJUST
53 ECG
RATE
ADJUST
Increases or decreases rate in 10%
increments; STD returns to standard
rate.
54
SIZE ADJUST
Increases and decreases ECG, BP, and
respiration waveforms or returns them to
standard size; use after selecting
desired channel.
55
AUTO TREND
Automatically varies NSR heart and
respiration rates and blood pressures;
Vtach episode 4 every 1 hour; trend
repeats every 2 hours.
DEFIB CONVERSATION
59 RHYTHM
SELECT
Permits selection of post defib rhythm,
which is then automatically inserted after
defib artifact.
60 DEFIB
DISABLE
Disables
auto defib sequence
RESPIRATION
71
APNEA
Thoracic impedance is simulated via
ECG output snaps.
72
RESP AT 10 BPM
Respiration waveform at 10 BPM
73
RESP AT 20 BPM
Respiration waveform at 20 BPM
74
RESP AT 40 BPM
Respiration waveform at 40 BPM
75
RESP AT 80 BPM
Respiration waveform at 80 BPM
76 CVA
COINCIDENCE
ECG and respiration are synchronized
and timed appropriately for
cardiovascular artifact. Impedance drops
immediately after each R wave.