Pac and pvc insertion – Fluke Biomedical DataSim 6100 User Manual
Page 29

Introduction and Specifications
Instrument
Familiarity
1
1-17
Table 1-5. Pendant Keypad Codes and Functions (cont.)
Code Function
Description
65
DAMPED LINE
Outputs a 50 mmHg flush waveform on
PA channel.
PA waveform follows flush. Trailing edge
of flush and PA pressure waveforms
demonstrate damped frequency
characteristics of pressure line.
66 CATHETER
WHIP
Pressure artifact caused by motion of
catheter; on PA channel
67 PRES/RESP
ARTIFACT
Varies level of respiration artifact on
pressures
PROCEDURES
68
IABP
Arterial pressure with cardiac assists
pressure pulse, 2:1 augmentation.
Simulation only; not interactive with the
IABP; output on arterial channel
69
PA WEDGE
Outputs a PAW waveform
70 SWAN-GANZ
INSERTION
Output on PA channel: right atrium to
right ventricle; right ventricle to
pulmonary artery; pulmonary artery to
wedged pulmonary artery; wedged
pulmonary artery to pulmonary artery
PAC and PVC Insertion
The third row of the keypad has three palpitation simulation keys:
•
9/PAC
•
PVC1
•
PVC2
A single PAC or PVC can be inserted into any rhythm by pressing one of these
keys. The respective PAC or PVC is inserted into the next cardiac cycle. Only
one PAC, PVC1, or PVC2 can be inserted at any one time. The eight-character
LCD display responds to key entries by displaying the abbreviated function
name.