Pac and pvc insertion – Fluke Biomedical DataSim 6100 User Manual
Page 29

Introduction and Specifications
Instrument
Familiarity
1
1-17
Table 1-5. Pendant Keypad Codes and Functions (cont.)
Code Function
Description
65
DAMPED LINE
Outputs a 50 mmHg flush waveform on 
PA channel. 
PA waveform follows flush. Trailing edge 
of flush and PA pressure waveforms 
demonstrate damped frequency 
characteristics of pressure line. 
66 CATHETER
WHIP
Pressure artifact caused by motion of 
catheter; on PA channel 
67 PRES/RESP
ARTIFACT
Varies level of respiration artifact on 
pressures 
PROCEDURES
68
IABP
Arterial pressure with cardiac assists 
pressure pulse, 2:1 augmentation. 
Simulation only; not interactive with the 
IABP; output on arterial channel 
69
PA WEDGE
Outputs a PAW waveform
70 SWAN-GANZ
INSERTION
Output on PA channel: right atrium to 
right ventricle; right ventricle to 
pulmonary artery; pulmonary artery to 
wedged pulmonary artery; wedged 
pulmonary artery to pulmonary artery 
PAC and PVC Insertion
The third row of the keypad has three palpitation simulation keys:
•
9/PAC
•
PVC1
•
PVC2
A single PAC or PVC can be inserted into any rhythm by pressing one of these 
keys. The respective PAC or PVC is inserted into the next cardiac cycle. Only 
one PAC, PVC1, or PVC2 can be inserted at any one time. The eight-character 
LCD display responds to key entries by displaying the abbreviated function 
name. 