Cardiac output personality modules – Fluke Biomedical DataSim 6100 User Manual
Page 54

DataSim 6100
Operators Manual
2-18
Cardiac Output Personality Modules
The Simulator can be used with either of two Cardiac Output (CO) Personality
Modules. 0 °C injectate temperature (model 6070-18) and 25 °C injectate
temperature (model 6070-20) both simulate the Edwards style blood
temperature catheter in generating several dilution curves.
The dilution curve at 0 °C injectate temperature assumes the following
conditions:
•
10 cc Injectate volume
•
Edwards 93-131-7F Catheter
•
Injectate Bath Probe, 90.22 k
Ω
(compatible with Switchcraft connector
SL-40-47F)
•
Computation Coefficient .542 *
The dilution curve at 25 °C injectate temperature assumes the following
conditions:
•
10 cc Injectate volume
•
Edwards 93-131-7F Catheter
•
Injectate Bath Probe 62.34 k
Ω
(compatible with Switchcraft connector
SL-40-47F)
•
Computation Coefficient .595 *
* Note
The use of other computation LM10 coefficients could result in CO
values other than specified.
To use the Cardiac Output Module:
1. Insert the Cardiac Output Module into the Simulator
Personality
Module
connector.
2. Power on the Simulator.
3. Connect the CO computer BT (blood temperature) cable to the CO
Module’s 3-pin connector. (The BT cable in the clinical setting is attached
to the thermistor on the Swan-Ganz catheter.)
4. Connect the CO computer injectate temperature cable to the black
injectate temperature box that simulates 0 °C or 25 °C temperature. *