Fluke Biomedical DataSim 6100 User Manual
Page 46

DataSim 6100
Operators Manual
2-10
5. To enable the automatic defib conversion feature and select the post defib
rhythm:
a. Press and release the
SPEC FUNC
key
b. Enter code number
59
to initiate the
DEFIB CONVERSION
Rhythm Select
function. The keypad displays:
SEL RTHM
c. Select the desired post-defib rhythm by pressing a Direct Function
key, or by pressing the
SPEC FUNC
key followed by the two-digit
code number of the desired rhythm.
Note
If a post defib rhythm is not programmed, the defib artifact is
followed by asystole until another rhythm is manually selected.
6. To initiate the defibrillation sequence:
•
Defibs equipped with apex/sternum paddles: Firmly place the paddles
on the electrodes located on the manikin chest skin or torso cover.
•
Defibs equipped with disposable electrodes: Plug in adapters and
connect cables from defib to adapter snaps.
7. Charge the defib to 25 joules for Arrhythmia Anne or up to 400 joules for
Chris Clean.
8. Press the
SIMULATE DEFIB
button.
W
Caution
To avoid
damage to the Simulator or adverse affects on its
performance, refer to
the Arrhythmia Anne and Chris Clean
instruction manuals for the proper procedures and use of
the defibrillator on the manikin. Under no circumstances
should a defibrillator be discharged directly into the
Simulator.
After defibrillation of the manikin, the Simulator simulates defibrillation
artifact, followed by the pre-selected rhythm. The programmed post
defibrillation rhythm can be changed at any time.