Fluke Biomedical DataSim 6100 User Manual
Page 32

DataSim 6100
Operators Manual
1-20
Waveforms ..........................................................Static, square wave, and physiological
dynamic
Output connector.................................................Switchcraft 15GM7F
Performance Testing
Static ...................................................................0 to 250 adjustable in 5 % increments
Square Wave ......................................................0 to 250 mmHg adjustable in 5 %
increments
Respiration
Output signals
Delta Impedance .................................................0.25 to 2.50 Ω Lead I
Base Impedance .................................................250 Ω Lead I, 750 Ω Lead II
Rate Range .........................................................0 to 80 BPM
Lead ....................................................................All leads
Performance Testing
Rate.....................................................................0, 10, 20, 40, and 80 BPM
Delta Impedance .................................................0.25, 0.50, 1.00, 1.50, 2.00, and
2.50 Ω
Coincidence check
General Information
Channels .............................................................ECG/arrhythmia, respiration, arterial
pressure, RA pressure, and auxiliary
channel for optional parameters.
Fundamental Rhythms and Sequences ..............Normal sinus rhythm
Sinus tachycardia and bradycardia
Ventricular and sinus asystole
Atrial tachycardia
Atrial flutter
Fibrillation
AV blocks (1st degree, 2nd degree
Mobitz I and II, and 3rd degree)
Unifocal and multifocal PVCs
Ventricular tachycardia
Ventricular fibrillation
PVCs at 1 to 35/min for any rhythm
Couplet
Triplet
Bigeminy
Junctional
Accelerated junctional
PACs and PJCs
Atrial tachycardia with aberrant
conduction
Idioventricular
Agonal
ST-segment elevation and depression
Pacemaker (atrial, ventricular, and AV
sequential)
Failure to capture and sense