DataSim 6100
Operators Manual
1-20
Waveforms ..........................................................Static, square wave, and physiological
dynamic
Output connector.................................................Switchcraft 15GM7F
Performance Testing
Static ...................................................................0 to 250 adjustable in 5 % increments 
Square Wave ......................................................0 to 250 mmHg adjustable in 5 % 
increments
Respiration
Output signals
Delta Impedance .................................................0.25 to 2.50 Ω Lead I 
Base Impedance .................................................250 Ω Lead I, 750 Ω Lead II 
Rate Range .........................................................0 to 80 BPM 
Lead ....................................................................All leads 
Performance Testing
Rate.....................................................................0, 10, 20, 40, and 80 BPM 
Delta Impedance .................................................0.25, 0.50, 1.00, 1.50, 2.00, and 
2.50 Ω
Coincidence check
General Information
Channels .............................................................ECG/arrhythmia, respiration, arterial
pressure, RA pressure, and auxiliary 
channel for optional parameters. 
Fundamental Rhythms and Sequences ..............Normal sinus rhythm
Sinus tachycardia and bradycardia 
Ventricular and sinus asystole 
Atrial tachycardia 
Atrial flutter 
Fibrillation 
AV blocks (1st degree, 2nd degree 
Mobitz I and II, and 3rd degree) 
Unifocal and multifocal PVCs 
Ventricular tachycardia 
Ventricular fibrillation 
PVCs at 1 to 35/min for any rhythm 
Couplet 
Triplet 
Bigeminy 
Junctional 
Accelerated junctional 
PACs and PJCs 
Atrial tachycardia with aberrant 
conduction 
Idioventricular 
Agonal 
ST-segment elevation and depression 
Pacemaker (atrial, ventricular, and AV 
sequential) 
Failure to capture and sense 