Respiration – Fluke Biomedical DataSim 6100 User Manual
Page 44

DataSim 6100
Operators Manual
2-8
To enable an artifact:
1. Using the keypad, press
SPEC FUNC
, followed by the numbers
47, 48,
or
49
. For example, using the number
47
, the keypad briefly displays:
60 CYCLE
followed by the prompt:
], [ or STD
2. Press
]
to increase or
STD
to reset to standard level. The
[
capability is
not available.
Table 2-5 shows the effect of artifact modification on the parameters.
Table 2-5. Artifact Modifier Effects
FUNC TYPE STD +1 +2 +3
47
ECG 60 Hz
0
.1
.4
.7 mV
48
ECG RESP
0
.4
.6
1.0 mV
49
ECG MUSCLE
0
.4
.6
1.0 mV
Respiration
The Simulator provides a varying thoracic impedance signal via all 10 ECG
electrode output snaps. Several respiration rates from apnea to 80 BPM can be
selected. After a respiration rate has been selected, the respiration rate remains
fixed during subsequent rhythm selections or until a new respiration rate is
selected.
To vary the amplitude of the thoracic impedance signal (R):
1. Using the keypad, press
SPEC FUNC
, followed by the numbers
54
. The
keypad briefly displays:
SIZE