Printing – Fluke Biomedical 2MF Index User Manual
Page 93
Appendices
Printing and Data Transfer
A
A-5
Printing
Sample Index 2MF Test Results Output
The Simulator Autosequence printout shown below requires using the previous settings
to establish communication between the the Simulator and a printer.
•
The optional header is all information from the “Control#” field down to the
“Technician” field.
•
Actuals are all information under the “Settings:”
heading. A normal report includes
only the fields Alarm Response Time, Rate Alarm Response Time, and Pulse
Amplitude Test.
Fluke Biomedical
SPO2 Simulator Ver: 2.00
--------------------------------------
Control#: __________________________
Serial #: __________________________
Model: _____________________________
Mfr: _______________________________
Location: __________________________
Technician: ________________________
Date: ______________________________
Simulations: NELLCOR
---------------------------------------
Settings:
Actual:
#
O2
RATE
O2
RATE
1
100%
60 BPM
99 59
2
95%
65 BPM
95 64
3
90%
70 BPM
90 70
4
85%
75 BPM
85 75
5
84%
80 BPM
84 80
6
83%
85 BPM
83 85
7
82%
90 BPM
82 90
8
81%
95 BPM
81 95
9
80% 100 BPM
80 100
10
79% 105 BPM
79 105
---------------------------------------
Alarm Response Time:12 Seconds at 95% O2
---------------------------------------
Hi Rate Alarm Response Time:10 secs at 80
BPM
Lo Rate Alarm Response Time:10 secs at 80
BPM
---------------------------------------
Pulse Amplitude Test
At 85%, the signal is lost.
---------------------------------------
Motion Response Time: 10 Seconds
---------------------------------------
End of test.
esl155.eps