Fluke Biomedical 2MF Index User Manual
Page 100

Index 2MF
Users Guide
D-2
Dynamic Test
Test Format........................................................... Photo detector/diode response to both the red and infrared light
generated by the probe when pulsed by an internal test signal.
Test Signal.......................................................... Pulsed between the two LEDs, constant current level at 1.0 mA
Test Results........................................................ Nominal range of 0 to 20,000
Checksum.............................................................. The sum of all locations in the program chip. For service only.
Note
The above specifications are for pre-programmed makes only, 20 % pulse
amplitude SpO2. Results for Presets, custom makes, and various pulse
amplitudes are unspecified and depend on the algorithm of the specific
pulse oximeter being tested.