Fluke Biomedical 2MF Index User Manual
Page 92

Index 2MF
Users Guide
A-4
4. Press R232 to set the RS232 port parameters:
esl167.eps
5. Press the softkey under the baud rate, parity type, and data and stop bit sizes to set the
correct communications parameters for your printer or other device.
Table A-1 describes parameters used for PC to peripheral communication.
Table A-1. Communication Parameters
Parameter Description
BAUD
Each press of the button decreases the baud rate from 9600 to 2400, 1200, 600,
and 300 bps. When 300 is displayed, a button push then shows 9600.
PAR
Pressing this button toggles the parity from No parity (N) to Even parity (E) to Odd
parity (O) and back.
DATA
Pressing this button toggles the data bits back and forth from 7 to 8.
STOP
Pressing this button toggles the stop bit back and forth from 1 to 2.
When you finish setting the communication control parameters:
•
Press esc to return to the Utilities menu.
•
Press esc
again to return to the Main menu 2.
•
Press more to return to the Main menu 1.