Setting and Changing Simulation Factors
Setting Simulations
6
6-3
Changing the Default SpO2 Setting
The default settings (for the current session from power-up, when no prior changes have
been made) are for a simulated patient with 96% saturated oxygen levels and with a pulse
of 75 beats per minute.
To change either of these settings, press MAN on the Simulations Menu.
fgg031.eps
The manual settings screen appears on the LCD, displaying the current SpO2 and BPM:
esl032.eps
Setting the O2 Saturation Level
You can raise or lower, as needed, the simulated degree of oxygen saturation in the
virtual patient.
1. To raise the O2 level two steps at a time, press
O2+
.
esl033.eps
2. To lower the O2 level two steps at a time, press
O2-
.
esl034.eps