Setting the r-value – Fluke Biomedical 2MF Index User Manual
Page 81

Manufacturers Mode
Accessing Manufacturers' Tests
11
11-5
The Waveform Source screen provides two options:
esl146.eps
2. Select the oximeter by pressing UUT (Unit Under Test) as the source for the pulse
amplitude signal.
esl147.eps
3. Alternatively, press DC. The Simulator will ignore the oximeter's pulse amplitude
and will use instead a fixed dc level for infrared and red pulse simulations.
esl148.eps
The signals will pass through the Simulator based on their original amplitude.
4. To return to the Advanced Options menu, press esc.
Setting the R-Value
You can also enter a custom R-value using the following procedure:
1. Access the Advanced Options menu (if you have not already done so) by pressing
OPTS on Main Menu 2.
2. Press RVAL on the Advanced Options menu:
esl149.eps
You will see a screen appear similar to the following: