Fluke Biomedical 2MF Index User Manual
Page 44

Index 2MF
Users Manual
7-2
esl049.eps
The display changes to present these options:
esl051.eps
When setting new oxygen limits, the Simulator displays the last setting used (or 96%
if not previously changed) and resets the time to 00.0 seconds.
2. Press the softkeys
+
or
-
to start the timer, and the screen displays the “Timing!”
message.
esl052.eps
•
Pressing the softkey
+
(plus) will increase by the step amount set previously
using the Simulations menu without affecting the timing function.
•
Pressing the softkey
–
(minus) will decrease by the step amount without
affecting the timing function.
Note
Continue pressing
+
or
-
until the O2 percent you want to test for is
displayed. Each press pf the softkey
+
or
–
resets the timer internally and
restarts the timing.
3. If the O2 setting is outside the oximeter's alarm limit, the alarm will sound on the
pulse oximeter being tested after a period of time. Press STOP immediately. This
stops the timer, and displays the “elapsed time to alarm” for the specified simulation
level.
esl053.eps
4. To print the results of the alarm, press PRT.