Fluke Biomedical 2MF Index User Manual
Page 110

Index 2MF
Users Guide
F-2
Attachment of the
FingerClip Probe
Detector Window
Properly Applied
Finger Stop
Attaching the Probe to Index
(Ohmeda FingerClip Probe)
Misapplied
esl157.eps
Figure F-1. Attaching the Probe to Index 2MF
The low readings may be because the Pulse Oximeter under test is a Nellcor unit but
the user has Ohmeda selected. Make sure the make selected in Index 2MF agrees
with the make of the pulse oximeter being tested. Many manufacturers use different
“R” curves for measurement.
You might try repositioning the probe for closer results, especially disposable probes.
With disposable probes, it is difficult to ensure consistent alignment of the LEDs to
the photo detector. This could also be a problem when using a disposable probe on a
patient.
5. I cannot find the make of the pulse oximeter I am testing with Index 2MF. What do I
do?
Index 2MF currently has curves for many oximeters. If the pulse oximeter under test
is one of the following, use the Nellcor make selection:
Airshields, Inc.
Ivy Biomedical
Atom Medical Corp.
Marquette (Eagle)
Baxter Healthcare Corp.
MDE(Medical Data Electronics)
Century Medical Inc.
Mennen Medical
Colin Medical Instruments Corp.
North American Drager
Corometrics Medical Systems
Pace-Tech. Inc.
Critikon, Inc.
Protocol Systems, Inc.
Datascope Corp.
Pryon Corp
Drager (Europe)
Siemens
Edentec Sleeptrace/Areca
Science
Fukuda Denshi
SpaceLabs
Synectics Witt
Biomedical