Using the optical finger probe (all versions) – Fluke Biomedical 2MF Index User Manual
Page 17
Introducing the Index 2MF
Getting Started
1
1-5
When the Main Menu 1 displays in the LCD, the following options are available:
esl002.eps
The Main Menu 1 consists of the primary test functions available to the Simulator, with
the softkeys
SIM, LMTS, PRBE, CUST
, and
more
. Press the up arrow below the
menu item to select a menu option, change a setting, or to advance to additional menus.
Note
If you see a message warning you that your battery is low, you need to
attach the battery charger. Call Fluke Biomedical for assistance if needed.
Toggle between Main Menu 1 and Main Menu 2 with the more softkey.
For example, you have these options at the Main Menu 1 screen:
Press the softkey SIM, and then MAN to open the Simulations menu. The menu
“Simulations:” appears in the Simulator LCD. This menu allows you to set oxygen
saturation levels and heart rate in beats-per-minute (BPM).
Press the softkey MAN to set the Sp02 levels from 35 to 100%, and the heart rate from 25
to 250 BPM.
1. The LCD indicates current Sp02 level and BPM. Additional softkey options now
available allow you to set the SpO2 levels (02+ and 02-), as well as increase and
decrease the BPM (BPM+ and BPM-).
Return to the previous menu, Main Menu 1, by pressing the esc softkey. For detailed
information on using the test features, refer to the relevant chapter in this users guide.
Using the Optical Finger Probe (All Versions)
Connect the finger probe of the pulse oximeter under test to the Simulator finger probe
attachment. Position the pulse oximeter LEDs on the bottom of the Simulator finger
probe attachment.
From the Main Menu 1, select the SpO2 and Heart Rate settings by pressing the
appropriate
+
or
-
(plus or minus) keys, and viewing the settings displayed on the the
Simulator LCD.
Note
Oximeters take from 5 to 20 pulses to respond to a change in simulated
SpO2 or BPM.