Fluke Biomedical 2MF Index User Manual
Page 39

Setting and Changing Simulation Factors
Using Preset Patient Conditions
6
6-5
Table 6-2. Level 0 Presets
Motion
Preset Name
Saturation
Pulse
Rate
Signal
Strength
Saturation Frequency Amplitude
[presets#]
Normal
98 %
55 BPM
5.0 %
-
-
0 %
00
Normal/ Tap
98 %
55 BPM
5.0 %
78 %
2.5 Hz
7 %
01
Normal/ Shiver
98 %
55 BPM
5.0 %
78 %
6.0 Hz
15 %
02
Weak Pulse
90 %
95 BPM
0.65 %
-
-
0 %
03
Weak Pulse/
Tap
90 %
95 BPM
0.65 %
60 %
4.3 Hz
1 %
04
Weak Pulse/
Shiver
90 %
95 BPM
0.65 %
60 %
6.0 Hz
3 %
05
Bradycardia
88 %
45 BPM
5.0 %
-
-
0 %
06
Brachycardia/
Shiver
88 %
45 BPM
5.0 %
68 %
6.0 Hz
10 %
07
Hypoxic
70 %
95 BPM
2.0 %
-
-
0 %
08
Hypoxic/ Tap
70 %
95 BPM
2.0 %
50 %
4.3 Hz
3 %
09
Hypoxic/
Shiver
70 %
95 BPM
2.0 %
50 %
6.0 Hz
8 %
10
Neonate
90 %
180 BPM
1.0 %
-
-
0 %
11
Neonate
/Shiver
90 %
180 BPM
1.0 %
70 %
6.0 Hz
5 %
12
Tachycardia
85 %
130 BPM
1.2 %
-
-
0%
13
Geriatric
92 %
95 BPM
2.4 %
-
-
0 %
14
Obese
93 %
90 BPM
3.0 %
-
-
0 %
15
Brady Tap #2
88 %
45 BPM
5.0 %
96/100 %
3.9 Hz
0/4 %
16
Hypox Tap #2
70 %
95 BPM
2.0 %
96/100 %
4.3 Hz
0/3 %
17
Weak Tap #2
80 %
95 BPM
0.9 %
96/100 %
3.6 Hz
0/1 %
18
Normal Tap #2
93 %
55 BPM
5.0 %
96/100 %
2.5 Hz
0/3 %
19
Asystole
91 %
90 BPM
2.0 %
96/98 %
1.1 Hz
0/1.1 %
20
LowFreq1
80 %
75 BPM
1.0 %
93/97 %
0.5 Hz
0/4.2 %
21
LowFreq2
70 %
75 BPM
1.0 %
96/100 %
0.5 Hz
0/4.2 %
22
Slow Tap
80 %
75 BPM
1.0 %
96/100 %
2.0 Hz
0/3.0 %
23
Note
When presets with tap or shiver are selected, there is a 10-second delay
before the motion simulation starts.This delay will prevent an oximeter
from "locking on" to patient parameters from a previous simulation.
The Simulator is preset to simulate the SpO2 and pulse rate ratios.
1. Access the Simulations menu by pressing the softkey SIM in Main Menu 1.