Testing the pulse oximeter limits, Introduction, Testing pulse oximeter limits – Fluke Biomedical 2MF Index User Manual
Page 43: Simulating oxygen conditions, Chapter 7

7-1
Chapter 7
Testing the Pulse Oximeter Limits
Introduction
Once you have determined the range in which you want to test a pulse oximeter for SpO2
and rate, you can test the pulse oximeter for upper and lower limits for response time.
Testing Pulse Oximeter Limits
You can set your oximeter to sound an alarm whenever any pre-determined limits are
reached during the testing of your pulse oximeter.
From the Main Menu 1, press the
LMTS
softkey.
fgg011.eps
The following menu appears in the LCD:
esl049.eps
Note
Exit this menu with the softkey
esc
, which always returns you to the
previous menu, in this case the Main Menu 1.
Simulating Oxygen Conditions
You can test the sensitivity of your pulse oximeter by setting the simulated oxygen levels.
1. To set the oxygen limits, press
LMTS
at the Main Menu 1 if you are not already
there. Then, press the softkey
O2
in the Limits menu.