Probe test selection, Led testing – Fluke Biomedical 2MF Index User Manual
Page 68

Index 2MF
Users Guide
9-2
Note
The MFE version ships with Nellcor and Ohmeda probe and oximeter
adapters. Please contact Fluke Biomedical regarding the availability of
additional adapter cables.
W
Caution
Attempting to create adapter cables without all details of the
oximeter probe-operating scheme could result in damage to the
oximeter, the Simulator, or both.
2. Press the
PRBE
softkey at Main Menu 1:
Fgg012.eps
The
PRBE
softkey initiates a testing sequence that includes:
•
Verifying that a Probe Test Board is installed in the Simulator
•
Checking whether or not a probe adapter cable is connected to the probe port
•
Performing all measurements for LED, photo detector, and resistance tests
esl125.eps
Probe Test Selection
The user can then choose the specific probe test.
•
LED and photodiode electrical test
•
Pin-to-pin resistances
•
Photodiode detector optical test
Note
During the photodiode test, the finger probe being tested should not be
attached to the Index finger.
LED Testing
The Probe test should end with the following menu displayed. Press the
LEDS
softkey: