Electrical Safety Analyzer

Detailed Specifications

81

Insulation resistance

Ranges ................................................................ 0.5 to 20.0 M

Ω

20.0 to 100.0 M

Ω

Accuracy

20 M

Ω

Range ..................................................

±

(2 % of reading + 0.2 M

Ω

)

100 M

Ω

Range ................................................

±

(7.5 % of reading + 0.2 M

Ω

)

Source test voltage ............................................. 500 or 250 V dc (+20 %, -0 %) 2.0

±

0.25 mA short-circuit current

Maximum load capacitance ................................. 1

μ

F

ECG Performance Waveforms

Accuracy

Frequency .......................................................

±

2 %

Amplitude ........................................................

±

5 % of 2 Hz square wave only, fixed @ 1 mV Lead II configuration

Waveforms

ECG Complex ................................................. 30, 60, 120, 180, and 240 BPM
Ventricular Fibrillation
Square wave (50 % duty cycle) ....................... 0.125 Hz and 2 Hz
Sine wave ........................................................ 10, 40, 50, 60, and 100 Hz
Triangle wave .................................................. 2 Hz
Pulse (63 ms pulse width) ............................... 30 BPM and 60 BPM