Electrical Safety Analyzer

How to Connect a DUT to the Product

11

The Product is intended for use with single-phase,
grounded power. It is not intended for dual, split-phase or
three-phase power configurations. It can be used with a
power system that supplies the correct voltages for
single-phase and is grounded, or is an isolated power
system.
Use the power cord for your country mains supply that is
not more than the voltage or power rating of the product.
Connect the cord into the power input connector and then
to the mains outlet.

How to Connect a DUT to the Product

You can connect a Device Under Test (DUT) a number of
different ways for a full electrical safety test. Figure
shows a DUT connected to the test receptacle, applied
parts posts, and a connection to the enclosure or
protective earth ground of the DUT.

How to Turn On the Product

Note

To make sure the high voltage indicator works,
look for it to illuminate at the power-up.

Push the power switch found on the left-side panel so the
“I” side of the ac power switch is down. The Product does
a series of self tests and then shows the message in
Figure when the self test has completed successfully.

gtv125.bmp

Figure 4. Product Ready for Operation