Equipment current test, Leakage current test – Fluke Biomedical ESA615 User Manual
Page 47
![background image](https://www.manualsdir.com/files/785315/content/doc047.png)
Electrical Safety Analyzer
How to Do Electrical Safety Tests
33
Equipment Current Test
To measure the current consumed by the DUT, push
. The Product shows the current that flows through
the mains connections of the test receptacle.
Leakage Current Test
The Product measures leakage current for different DUT
configurations. The Product measures leakage found on
the enclosure and the earth connection as well as leakage
on each connected applied part and combinations of
connected applied parts.
The available leakage tests are set by the standard you
set in setup. See the How to Set the Test Standard section
to change the test standard.
Table is a list of six leakage current tests. Their names
are different when the standard is changed in the Product.
Push to access the leakage current main menu
shown in Figure .
Table 5. Test Names Based on Selected Standard
IEC60601 AAMI/NFPA
99
Protective Earth Resistance
Ground Wire Resistance
Earth Leakage Current
Ground Wire Leakage Current
Touch or Enclosure Leakage Current
Chassis Leakage Current
Patient Leakage Current
Lead to Ground Leakage Current
Patient Auxiliary Leakage Current
Lead to Lead Leakage Current
Mains on Applied Part (MAP) Leakage Current
Isolation Leakage Current