Table 7. Test Settings for Test Sequences

Test Setting

Description

Default

Value

Pause after power on

If set to yes, delays the start of the test by the time set in the Power on delay
parameter when power is applied to the DUT.

No

Pause before power off If set to yes, delays the start of the test by the time set in the Power off delay

parameter when power is removed from the DUT.

No

Power on delay

The time the Product will wait before it does the subsequent step after power is
applied to the DUT. The range is 0 to 9999 seconds.

2 sec

Power off delay

The time the Product will wait before it does the subsequent step after power is
removed from the DUT. The range is 0 to 9999 seconds.

0 sec

Test Speed

[1]

When set to Normal, the Product does a leakage current measurement in
5 seconds and an insulation resistance test in 1 minute.
When set to Rapid, the Product measures leakage current as fast as possible and
an insulation resistance test in 3 seconds.

Normal

Test Mode

When set to Automatic, the Product does each test sequence step automatically.
When set to Step-by-Step, you must push the Next Step softkey to move to the
subsequent step in the sequence.

Automatic

Halt on Test Failure

When set to Yes, the Product stops the test sequence when a fault is sensed.

Yes