Fluke Biomedical ESA615 User Manual
Page 82
![background image](https://www.manualsdir.com/files/785315/content/doc082.png)
ESA615
Users Manual
68
Table 7. Test Settings for Test Sequences
Test Setting
Description
Default
Value
Pause after power on
If set to yes, delays the start of the test by the time set in the Power on delay
parameter when power is applied to the DUT.
No
Pause before power off If set to yes, delays the start of the test by the time set in the Power off delay
parameter when power is removed from the DUT.
No
Power on delay
The time the Product will wait before it does the subsequent step after power is
applied to the DUT. The range is 0 to 9999 seconds.
2 sec
Power off delay
The time the Product will wait before it does the subsequent step after power is
removed from the DUT. The range is 0 to 9999 seconds.
0 sec
Test Speed
[1]
When set to Normal, the Product does a leakage current measurement in
5 seconds and an insulation resistance test in 1 minute.
When set to Rapid, the Product measures leakage current as fast as possible and
an insulation resistance test in 3 seconds.
Normal
Test Mode
When set to Automatic, the Product does each test sequence step automatically.
When set to Step-by-Step, you must push the Next Step softkey to move to the
subsequent step in the sequence.
Automatic
Halt on Test Failure
When set to Yes, the Product stops the test sequence when a fault is sensed.
Yes