Introduction, Electrical safety analyzer – Fluke Biomedical ESA615 User Manual
Page 15
![background image](https://www.manualsdir.com/files/785315/content/doc015.png)
1
Electrical Safety Analyzer
Introduction
Warning
To prevent possible electrical shock, fire, or
personal injury, read all “safety information”
before you use the Product.
The Fluke Biomedical ESA615 Electrical Safety Analyzer
(the Product) is a full-featured, compact, portable
analyzer, designed to verify the electrical safety of
medical devices. The Product tests to domestic
(ANSI/AAMI ES1, NFPA 99) and international (IEC62353,
AN/NZS 3551, and parts of IEC 60601-1) electrical-safety
standards. The Product simulates ECG to do
performance tests on ECG monitors.
The integrated ANSI/AAMI ES1 and IEC60601-1 patient
loads are easily selectable.
The Product does these tests:
•
Line (Mains) voltage
•
Ground Wire (Protective Earth) resistance
•
Equipment
current
•
Insulation
resistance
•
Ground (Earth) leakage
•
Chassis (Enclosure) leakage
•
Lead to Ground (Patient) and Lead to Lead (Patient
Auxiliary) leakage
•
Lead isolation (Mains on applied parts leakage)
•
Differential
leakage