Set the beeper, Set the display contrast, How to view instrument information – Fluke Biomedical ESA615 User Manual

Page 34: How to do electrical safety tests

background image

ESA615
Users Manual

20

Set the Beeper

To enable or disable the beeper:
1. From the setup menu, push the Instrument Setup

softkey.

2. Push

or  until the Beeper variable is

highlighted.

3. Push

.

4. Push

or  to highlight Off or On.

5. Push

.

Set the Display Contrast

There are two procedures to set the display contrast.
From the Test Sequence start-up menu or through the
setup menu.
When the Product shows its start-up menu (Select a
test…), push  or  to increase or decrease the display
contrast respectively. Push the Done softkey to exit
contrast setup.
To adjust the contrast through the setup menu:
1. From the setup menu, push the Instrument Setup

softkey.

2. Push

the

Display Contrast softkey.

3. Push

or  to increase or decrease the display

contrast respectively.

4. Push

the

Done softkey to exit contrast setup.

How to View Instrument Information

To show model number, serial number, firmware version,
and last calibration date of the Product, push . Next
push the Instrument Information softkey.

How to Do Electrical Safety Tests

The Product does a number of different electrical and
performance tests on biomedical equipment. The sections
that follow are test descriptions with instructions on how
to do them with the Product.