ESA615
Users Manual

74

F1 - F2

gtv111.eps

Figure 40. Fuse Access

5. Use a multimeter to measure the continuity of each

fuse.
If a fuse does not show continuity, replace the fuse(s)
with a fuse of the same current and voltage rating.
Applicable fuse ratings are posted on the case bottom
label of the Product. Table is a list of available fuses
with Fluke Biomedical part numbers.

6. Reinstall the fuse door and secure it with the screw.

How to Clean the Product

 Warning

To prevent electric shock, do not clean the
Product plugged into mains or attached to a
DUT.

Caution

Do not pour fluid onto the Product surface.
Fluid in the electrical circuitry can cause the
Product to fail.

Caution

Do not use spray cleaners on the Product.
This can cause fluid to leak into the Product
and damage electronic components.

Clean the Product occasionally with a damp cloth and
weak detergent. Try to prevent the entrance of liquids.
Clean the adapter cables with the same precautions.
Examine them for damage and deterioration of the
insulation. Examine the connections for integrity before
each use.