Maintenance

73

Delete a Set of Test Results

To delete a test result:
1. Push

.

2. Push

or  to highlight a test result name.

Note

Push softkey F1 to toggle between sort results by
date or sort A-Z.

3. Push

the

Delete softkey.

A delete confirmation screen will show in the display.

4. Push

the

Delete softkey.

Maintenance

 Warning

To prevent possible electrical shock, fire, or
personal injury:

Have an approved technician repair the
Product.

Use only specified replacement fuses.

Use only specified replacement parts.

Remove the input signals before you
clean the Product.

Disconnect the mains power cord before
you remove the Product covers.

Turn off the Product and remove the
mains power cord. Stop for two minutes
to let the power assemblies discharge
before you open the fuse door.

Do not operate the Product with covers
removed or the case open. Hazardous
voltage exposure is possible.

The Product is a calibrated measurement instrument. Use
the necessary precautions to prevent mechanical abuse
that could change the calibrated adjustments.

Fuse Test and Fuse Replacement

 Warning

To prevent electric shock, remove all power
cords and test leads from the Product before
you open the fuse door.

For electrical protection of the equipment outlet, the
Product uses two fuses, one in the live (L1) line and one in
the neutral (L2) line. To do a fuse test:
1. Turn the Product so the case bottom is up. See

Figure .

2. Flip up the tilt stand.
3. Remove the screw in the fuse door with a #2 Phillips

head screwdriver and lift the fuse door from the
Product.

4. Remove the two fuses from the Product.