How to make a test sequence, Make a new test sequence – Fluke Biomedical ESA615 User Manual
Page 80
ESA615
Users Manual
66
How to Make a Test Sequence
You can make a new test sequence from a different test
sequence or make a new one.
Make a New Test Sequence
To make a new test:
1. Push
.
2. Push
the
Test Library softkey.
3. Push
the
New softkey.
A test sequence wizard steps you through the
configuration of the test sequence. There are five setup
steps.
1. Push
or to highlight a test standard and push
.
2. Push
or to highlight a device class and push
.
3. Configure the Applied Parts (A.P.).
Note
A USB keyboard or barcode reader can be used
to type in the A.P. name.
If the DUT does not have applied parts, push the Next
Step softkey to skip the applied parts configuration.
If the DUT has applied parts, push New A.P.. You
must name the applied part and show how many DUT
applied parts are in the configuration.
The Product has five applied parts inputs. If you set
more than five applied parts, an error message shows
in the display. If the DUT has more than five applied
parts, you can use the 1-to-10 adapter. See the How
to Use the 1-to-10 Adapter section to learn more.
If you attach multiple applied parts to one input, then
you must set the Tied Setting to Tied. The Product
sets the subsequent available input for the applied
parts that are tied together. The position icon in the
display shows inputs that are configured as dots and
inputs not configures as circles. Figure shows the
position icon for one input configured and four inputs
not configured.
gtv127.bmp
Figure 37. Input Position ICON
To put a DUT applied part on each of the Product
inputs, set Tied Settings to Not Tied. The Product
will show the subsequent available inputs as
configured for this configuration.
The last applied parts variable is the Type. Highlight
the Type variable and push . Highlight a type
from the type list and push .