Do a Test Sequence
To do a test sequence:
1. Push
to show the test sequence screen in
Figure 38.
gtv125.bmp
Figure 38. Test Sequence Screen
2. Push
the
Test Library softkey.
3. Push
or to highlight a test sequence name.
Note
Push softkey F1 to toggle between an A to Z, Z to
A, by date sort of the test sequences.
4. Push
.
Note
To see the details of the test sequence you are
about to start, push the Sequence Details
softkey.
5. Push
the
Next softkey.
The equipment information screen in Figure 39 shows
in the display.
gtv126.bmp
Figure 39. Equipment Information Screen
The number, serial number, manufacturer, model, and
location of the Device Under Test can be typed in to help
identify the DUT on the test results.