Fluke Biomedical ESA615 User Manual
Page 81
![background image](https://www.manualsdir.com/files/785315/content/doc081.png)
Electrical Safety Analyzer
Test Sequences
67
Push the Done softkey to complete the applied part
configuration.
To edit an applied part configuration, use or to
highlight the A.P. configuration and push .
4. Edit test settings.
Table 7 is a list of test settings with their descriptions
and default values.
To set one of the test setting, push or to
highlight a test setting and push . After you
change the setting, push the Done softkey.
5. Name the test sequence. When you push the Next
Step softkey in the edit test settings step, the Product
automatically names the sequence Test Sequence
followed by the date and time. To accept the default
name, push the Next Step softkey. To change the
name, push the Edit softkey.
Note
A USB keyboard or barcode reader can be used
to edit the test sequence name.
The test sequence is complete. To put the test sequence
on the memory card, push the Save softkey. To see the
test sequence parameters before you save it, push the
Edit softkey.