Electrical Safety Analyzer
Test Sequences
69
Table 7. Test Settings for Test Sequences (cont.)
Test Setting
Description
Default
Value
Multiple PE Tests
If set to Yes, the Product prompts you to repeat PE tests or continue.
No
Multiple Non-Earth
Leakage
If set to Yes, the Product prompts you to repeat non-earth leakage tests or
continue.
No
Patient Lead Records
storage
[2]
When set to Store all, all results are put into the test results. If set to Store
worst/last, only the worst measured value is put into the test results. If set to store
worst/last and all measured values are the same, then the last value is put into
the test results.
Store
worst/last
Insulation Test Voltage Sets the insulation test voltage to 250 V dc or 500 V dc.
500 V dc
Reverse Polarity
If set to No, reverse polarity tests will be skipped.
Yes
[1]
You can change the test speed parameter on a test sequence to decrease the time necessary to do a test sequence.
[2]
Applies to leakage current and multiple PE tests only.