Electrical Safety Analyzer

How to Do Electrical Safety Tests

21

Set the Test Standard

The electrical safety tests in the Product are specified by
different safety standards: AAMI ES1/NFPA99,
IEC62353, IEC60601-1, and AN/NZS 3551. AAMI is set
as the default standard. To select different standard:
1. From the setup menu, push  until the Standard

variable is highlighted.

2. Push

.

3. Push

or  to highlight one of the standards.

4. Push

.

Not all electrical tests are applicable for all standards. In
these cases, the menu only shows the tests that are
specified in the set standard.

Note

The standard set in the setup menu applies to all
manual measurement modes. The automated
tests use the standard set for the selected test
sequence.

Mains Voltage Test

The Mains Voltage test measures the voltage on the
mains input through three measurements. To access the
Mains Voltage test, push . The Mains Voltage test
menu is shown in Figure .

gtv104.bmp

Figure 10. Mains Voltage Test Menu

Push each function softkey to do each of the three
measurements: Live to neutral, neutral to ground, and live
to ground.

Note

Power is removed from the test receptacle while
the Product does a Mains Voltage test.