Setting the flow control, Resetting to default values – Avery Dennison Monarch 9855 Operator Handbook User Manual
Page 60

5-4 Operator's Handbook
S e t t i n g t h e F l o w Co n t ro l
To change the setting, from the Main Menu select Setup, Port Settings, then
Serial Comm. Then, follow these steps.
1. Press or until you see
SERIAL COMM
Flow Control
2. Press Enter/Pause. The current setting is displayed, for example:
FLOW CONTROL
DTR
3. Press or to display the flow control you need, then press
Enter/Pause. Press Escape/Clear until you see the Main Menu.
XON is 17; XOFF is 19. Set flow control to DTR for PC computers
(unless you have XON/XOFF software).
R e s e t t i n g t o D ef a u l t V a l u e s
To change the setting, from the Main Menu select Setup, Port Settings, then
Serial Comm. Then, follow these steps.
1. Press until you see
SERIAL COMM
Reset
2. Press Enter/Pause.
Are you sure?
No
Press
or to make your selection, then press Enter/Pause. If you select
"Yes," the following defaults are restored: 9600 baud, No parity, 8 bit word
length, and 1 stop bit. Press Escape/Clear until you see the Main Menu.