3.14 Standard Measurements

Standard measurements are always available through the MEAS key on the front panel. These
measurements are spread across two to four screens to enhance readability. Switching between
these screens can be done by successively pressing the MEAS button on the front panel. This
will cause the screen to cycle through all available measurement screens.

3.14.1 CSW Series Measurements

For CSW Series, the following four measurement screens are available:

Mode

AC

DC

AC+DC

MEASUREMENTS 1

VOLTAGE

AC rms voltage

DC Voltage

AC rms voltage

CURRENT

AC rms current

DC Current

AC rms current

FREQUENCY

Frequency

n/a

Frequency

POWER

Real power

Power

n/a

MEASUREMENTS 2

VA POWER

Apparent power

power

Apparent power

PEAK CURR

Highest AC current

found

Highest DC current

found

Highest AC

current found

POWER FACT

Power factor

n/a

Power factor

CREST FACT

Crest factor

n/a

Crest factor

MEASUREMENTS 3

VOLT THD

Voltage distortion

n/a

Voltage distortion

CURR THD

Current distortion

n/a

Current distortion

INST PK CURR

Instantaneous peak

current

Highest DC current

found

Instantaneous

peak current

PHASE

Phase angle

n/a

Phase angle

The CSW Series has a fourth measurement screen for harmonics and trace analysis
measurements. This subject is covered in the next chapter.

Measurements are always running in the background. When the user selects a measurement
screen for display, the AC source first updates all the measurement parameters before
displaying the requested screen. This process may take up to a second. Consequently, pressing
the MEAS key may not always bring up the selected screen immediately. There will be a
perceptible delay. This will prevent the screen from appearing with invalid or blank readouts.

The measurement method for voltage and current will depend on the power source operating
mode. The following table shows the return value type (rms or average) and method of coupling
when the measurement command is initiated with a different extension at various operating
modes (AC, DC or AC + DC).

Measurement
Extension
and Coupling

Operating Mode

AC

DC

AC + DC

AC

rms

rms

rms

DC

rms

rms

average

Coupling

AC

DC

DC