AMETEK CSW Series User Manual
Page 83

User Manual
California Instruments
CSW Series
83
CONFIGURATION SETUP 3
MS704
Indicates the presence of the MIL/STD-704 Revision A through
F test option. If this option is installed, this field will show ON. If
this option is not installed, this field will show N/A (not
available). This field is available on Series II i/CSW systems
only.
ABD
Indicates the presence of the ABD0100.1.8 test option. If this
option is installed, this field will show ON. If this option is not
installed, this field will show N/A (not available).
LF
Indicates the presence of the Low Frequency limit option. If this
option is set, the maximum frequency that can be programmed
is 500 Hz.
SYSTEM
This field sets the controller for the correct multi box system
model configuration.
If this field is set incorrectly, the current limit scaling and current
measurement will be off by a factor of three. This field is
protected and cannot be changed unless the optional MB
configuration has been enabled.
NOTE: Options –A350, AMD24 and –B787 are only visible using the *OPT? bus command. Refer to
CONFIGURATION SETUP 4 screen.
CONFIGURATION SETUP 4
Note that the fourth configuration screen is only available on models with firmware revision 4.60
or higher. For units with older models, refer to the CONFIGURATION SETUP 3 screen.
MB
This configuration parameter does not exist in the CSW series.
MANUAL
This configuration parameter does not exist in the CSW series.
OPT0, OPT1, OPT2
This field indicates the presence of the following three options:
OPT0 = A350 Avionics option (-A350)
OPT1 = AMD24 Avionics option (-AMD)
OPT2 = B787 Avionics option (-B787)
The scroll through the three available OPTn fields, use the
shuttle. ON indicates the option is present, N/A indicates the
option is not installed.
LAN
This field indicates the presence of the LAN Ethernet interface
option. ON indicates the option is present, N/A indicates the
option is not installed.