AMETEK CSW Series User Manual
Page 130
User Manual
California Instruments
130
CSW Series
VOLT FREQ MAX
This test will set the voltage and frequency to levels defined by Table 9-2. The test will last for 30
minutes. The test will be repeated for the EURO standard using the Voltage setting from
Table 9-1 and the frequency from Table 9-2. The CLR Key in local operation will terminate the test
at any time. Group execute trigger will terminate the test remotely. The unselected phases will
remain at 115 volts. The
←
key (backspace) will terminate the test at any time.
VOLT UNBALANCE
Standard/Group
RTCA
Group1
Group2
Group3
Voltage offset
6
6
6
9
Frequency
400
400
360/650
360/800
Table 8-3: Normal Voltage Unbalance
This test will change the output voltage for each phase from 115 volts to 115V + offset. Refer to
Table 8-3 for the offset value and the Frequency. The test will last 30 minutes. The test will be
repeated for a second Frequency if applicable. The test can be terminated at any time.
The
←
key will terminate the test at any time.
WAVEFORM DISTORTION
This test will generate a 5% THD voltage distortion on the output voltage waveform. The
distortion is generated by using a clipped sine wave. The test will last for 30 minutes. The
←
key (backspace) will terminate the test at any time.
VOLTAGE MODULATION
This test requires a numeric value entry equal to the modulation rate in Hz. This entry value
must be between 1 Hz and 200 Hz. The amplitude modulation is calculated based on the
modulation rate as defined in Figure 8-4. This test will last for 2 minutes.