PHASE C

Phase angle of phase C with respect to phase A in three phase
mode. If the AC source is a single phase model, this field will
shown 0°. If the AC source is a split phase model, this field will
shown 180°.

3.11.11.5 CONFIGURATION SETUP screens

Figure 3-31: CONFIGURATION SETUP Menus

The configuration setup screen is not a menu but only serves to inform the user of the software
options installed in the AC source. The cursor can be moved to any of the fields in this screen
but none of these fields can be changed. The following information is provided on this screen:

Entry

Description

CONFIGURATION SETUP 1

NO. OUTPUT

Displays the phase mode option. SELECT indicates the phase
mode option is installed and the user can select between single
and three phase modes of operation.

FIXED indicates the phase mode option is not installed and
only single phase or three phase mode of operation is possible.

ADVANCE

This field indicates the presence of advanced capabilities for
waveform generation and measurement analysis are present.

DO160

Indicates the presence of the RTCA DO160 test option. If this
option is installed, this field will show ON. If this option is not
installed, this field will show N/A (not available).

MIL704

Indicates the presence of the MIL/STD-704 Revision D and E
test option. If this option is installed, this field will show ON. If
this option is not installed, this field will show N/A (not
available).

CONFIGURATION SETUP 2

IEC 4-11

Indicates the presence of the IEC 61000-4-11 test option. If this
option is installed, this field will show ON. If this option is not
installed, this field will show N/A (not available).

IEC 4-13

Indicates the presence of the IEC 61000-4-13 test option. If this
option is installed, this field will show ON. If this option is not
installed, this field will show N/A (not available).

CLOCK/LOCK

Indicates the presence of the -LKS clock and lock auxiliary
option. For units without -LKS, this field is set to N/A. For units
with the -LKS option installed, this field is set to ON. Note that
master unit (with -LKM) also has this field set to N/A.

WH METER

Indicates the presence of the Watt Hour Meter option.