2.1.3 Measurements

(All specifications are at 25˚C unless noted otherwise)

Parameter

Accuracy (

±

)

Resolution

Frequency

2 counts
0 to 45

˚C

0.01: 2.00 to 81.91 Hz
0.1: 82.0 to 819.0 Hz
1.0: 820 to 7990 Hz

RMS Voltage

±0.1% of range from 5 to 156 or 10
to 312 volts. Above 1kHz add
0.2%/kHz

0.01 Volt

RMS Current

±1% of range add ±1.5%/kHz above
500 Hz

Ranges: 0.5 to 16A: 156V range

0.5 to 8A: 312V range

Multiply by 3 for 1-phase mode

0.001 Amp

Peak Current

±5% of range, 40 to 500 Hz; add
±1%/kHz 500 to 5 kHz

Ranges: 0 to 56A; 156V range

0 to 28A; 312V range

Multiply by 3 for 1-phase mode

0.01 Amp

VA Power

±2.5% of range, DC or 40 to 500.
Add ±1%/kHz above 500 Hz.

Ranges: 1.8kVA; 3-phase mode

5.6kVA; 1-phase mode

1 VA

Real Power

±2.5% of range, DC or 40 to 500.
Add ±1%/kHz above 500 Hz.

Ranges: 1.8kW; 3-phase mode

5.6kW; 1-phase mode

1 W

Power Factor
(>0.2kVA)

±5% of range at full power, 40 to
500 Hz. Add ±1%/kHz above 500 Hz

0.01

Temperature Coefficient

<±50 ppm for all functions above

Phase

±2˚, 40 to 500 Hz, add ±2˚/kHz
above 500 Hz. (0 to 45

˚C)

±1˚ for outputs above 20 VRMS

Current and Power Accuracy specifications are percent of Range. The Ranges are listed below .