Calorimeter Instruction Manual
5-14
Moisture Multiplier:
Press to access the Moisture Multiplier numeric dialog box and set the value. This value is
associated with the net heat calculation. It is related to the heat of vaporization of water.
Data Entry Controls Menu
Prompt for Bomb ID:
Toggles ON or OFF. In the ON
position the controller will prompt for
a Bomb ID (1-4) when a test is
started.
Weight Entry Mode:
This key steps through the options
for entering sample weights either
manually through the touch screen,
balance port or through the network.
Acid Entry Mode:
This key steps through the options
for entering acid correction value either manually through the touch screen or
automatically through the balance port.
Hydrogen Entry Mode:
This key steps through the options for entering hydrogen content for calculating the net
heat of combustion either manually through the touch screen or automatically through
the balance port.
Auto Sample ID Controls:
Accesses sub-menu for controlling the automatic assignment of sample identification
numbers.
Automatic Sample ID:
When set to ON the unit will
automatically assign sample
identification numbers in
accordance with parameters set
by the other three keys on this
menu. When set to OFF, the
user manually enters each
sample ID when prompted to do
so.
Auto Sample ID Prefix:
An entry here will be used as a
prefix for all sample IDs, if the
Automatic Sample ID is set to
ON. Press this key to access a
sub-menu for entering an alphanumeric prefix.